BIOTECHNOLOGIA ACTA, V. 8, No 6, 2015
UDK 543.6 + 577.15 + 543.555 + 544.475
doi: 10.15407/biotech8.06.016
INVESTIGATION AND OPTIMIZATION OF REACTIVATION OF UREASE BIOSENSOR FOR HEAVY METALS INHIBITION ANALYSIS
Budnikov S. R.1, 2 Soldatkin O. O.1, 2 Kukla A. L.3 Khomenko 1.1.1, 4
Dzyadevych S. V.1, 2 Soldatkin O. P. 1, 2
institute of Molecular Biology and Genetics of the National Academy of Sciences of Ukraine, Kyiv 2Taras Shevchenko National University of Kyiv, Ukraine 3 Lashkaryov Institute of Semiconductor Physics of the National Academy of Sciences of Ukraine, Kyiv 4Centre of Intellectual Property and Technology Transfer of the National Academy of Sciences of Ukraine, Kyiv
E-mail: [email protected]
Received 14.09.2015
Research was aimed at the optimization of the urease biosensor for analysis of heavy metals and determining the opportunities of its reactivation. A differential pair of gold interdigitated electrodes deposited on a ceramic substrate was used as the conductometric transducer. As a bioselective membrane served urease, coimmobilized with bovine serum albumin by glutaraldehyde cross-linking on the surface of conductometric transducer. 1.0 mM urea was selected as an optimal substrate concentration for the inhibition analysis of heavy metals. The biosensor was tested for its sensitivity to different heavy metals, the calibration curves were plotted. The proposed biosensor was shown to have high reproducibility of signals prior and after inhibition, the measurement error was less than 3%. It was proved a possibility of reactivation of the bioselective membrane after urease irreversible inhibition by heavy metals, using the ethylenediamine tetraacetic acid solution. The optimum conditions of reactivation, i.e. the dependence of its level on time and concentration of heavy metals, were determined. The additional stage, post-inhibition reactivation, was shown to increase significantly the selectivity of biosensor determination of heavy metals.
Key words: conductometric transducer, biosensor, urease, heavy metals, inhibition analysis, enzyme reactivation.
At present, one of the most acute environmental problems is the contamination with waste products [1]. Among the various pollutants, heavy metals and their compounds are notable by abundance, high toxicity, ability to accumulate in living organisms [2]. They are widely used in manufacture, therefore, despite the clean-up, the content of heavy metals in industrial wastewater is rather high.
They also enter the environment with sewage, smoke and dust from industrial enterprises. Many metals form stable organic compounds, high solubility of which facilitates the migration of heavy metals in natural waters. More than 40 chemical elements are referred to as heavy metals, but significantly fewer number should be controlled taking into account toxicity, persistence, abundance, an ability to accumulate in the environment [3, 4].
In the aquatic environment, microparticles of heavy metals are present as solutions,
suspensions and colloids. Free hydrated ions, simple inorganic and organic complexes are soluble [5-9]. Currently, there are many methods, which are widely used for the determination of heavy metals, such as atomic absorption spectrometry, plasma mass spectrometry, etc. However, these methods utilize sophisticated appliances and are unsuitable in the field [10]. Electrochemical biosensor methods in general are more reasonable in the field because they do not require complex equipment [11-13].
Biosensor determination of heavy metal ions is based on their inhibition effect on enzymes. For example, ions of copper, cadmium, mercury, zinc are effective inhibitors of the urease activity in the urea hydrolysis. An important stage in the running of urease biosensor is a procedure of its reactivation after inhibition by heavy metal ions. An exposure of the inhibited biosensor in
16
Experimental articles
buffer solution for a long time does not restore the enzymatic activity of its biomembranes whereas EDTA solution acts as efficient reactivator [14].
Materials and Methods
Materials
In the work, the enzyme urease from soya beans with activity of 66 U/mg was used (Fluka, Germany). Bovine serum albumin (BSA, fraction V), 50% aqueous solution of glutaraldehyde (GA) and urea were from Sigma-Aldrich Chemie (Germany).
The compounds for the preparation of buffers, inhibitors, reactivator and other inorganic compounds used in the work were of domestic production and had a chemical purity grade.
Conductometric transducers
Conductometric transducers were produced in Lashkaryov Institute of Semiconductor Physics (Kyiv, Ukraine) in accordance with our recommendations. They consist of the sital substrate (fused Al2O3) 5 x 30 mm in size with a pair of gold interdigitated electrodes. The 0.1 ц thick titanium sublayer is used for better adhesion of metal to the substrate. More information on the transducers used is in the previous paper [15].
Scheme of experimental setup for conductometric measurements
The changes in conductivity of nearelectrode layer of conductometric transducer were determined by the conductometric measuring unit. The differential mode of measurement was used to increase the sensor sensitivity and minimize noise caused by nonspecific effects. The scheme and function of this conductometric setup have been described earlier [16, 17].
Preparation of bioselective elements
To prepare working urease-based bioselective membranes, the solution containing 5% of urease, 5% of BSA and 10% of glycerol in 20 mM phosphate buffer, pH 6.5, was used. The mixture for the reference membrane was prepared in the same way, but no enzyme was taken, only BSA of the final concentration equivalent to the amount of protein in the membrane, i.e. 10%. The solutions prepared were deposited onto the transducer working parts by micropipettes to cover completely the working electrode surface. The transducers with deposited membranes were incubated in saturated
glutaraldehyde (GA) vapor for 25 min and dried for 5 min in air at room temperature. Then the biosensors were washed for 6 min in the working buffer to remove excess unbound GA and other components of the membrane (changing the buffer every 2 min).
Methods of measurement
Measurements were carried out at room temperature in an open cell filled with 5 mM phosphate buffer, pH 6.5, with constant stirring. The specified substrate concentration in the working cell was obtained by adding the aliquots of substrates stock solutions. For inhibition analysis, the responses to substrate were evaluated before inhibition, then the biosensor was incubated for 20 min in a solution of heavy metal ions of different concentrations and washed from excess inhibitor, and the response to substrate was re-assessed. In this way, a level of the enzyme inhibition, which corresponds to the concentration of heavy metal ions in the sample, was determined. Reactivation of the inhibited bioselective elements was performed by incubation in 10 mM EDTA solution for 30 min. Then they were washed from excess reactivator, and the response to the substrate was measured again.
All experiments were conducted in 3-5 series. Non-specific changes in the output signal associated with electrical noise and fluctuations in environmental temperature and pH were avoided due to use of a differential mode of measurements. In the statistical analysis of the results obtained, the values of arithmetic mean and its standard deviation were calculated; the data were considered significant at P < 0.05.
Results and Discussion
Study of main analytical characteristics of biosensor in direct substrate determination
The biosensor operation is underlain by the enzymatic reaction in the membrane containing urease deposited on the surface of conductometric transducer:
Urease
H2N-CO-NH2 + 2H2O +H+ ^ 2NH4+ + HCO3-.
In the course of the reaction, protons are absorbed, which leads to changes in pH and generation of additional ions (NH4+ and HCO3) in the working membrane [18]. This changes the solution conductivity, which is registered by the conductometric transducer.
17
BIOTECHNOLOGIA ACTA, V. 8, No 6, 2015
In inhibition analysis, first it was necessary to determine an optimal concentration of urea as a substrate, i.e. to choose the urea concentration, at which the biosensor sensitivity to heavy metal ions is maximal. It is known that at irreversible inhibition the largest biosensor sensitivity to toxins can be reached if the substrate concentrations are within the range of biosensor saturation with the substrate when each enzyme molecule is involved in the conversion of this substrate. The graph of dependence of biosensor responses on the urea concentration in solution (Fig. 1) shows that above 1.0 mM complete biosensor saturation with the substrate is observed. Therefore, in further experiments on inhibition analysis 1.0 mM substrate was used.
It was necessary to confirm that a decrease in the biosensor response to the substrate after its incubation in a solution of heavy metals occurs due to the inhibition of bioselective element, but not because of the measurement error. Therefore, at the next stage of work the biosensor operational stability and reproducibility of its signal were tested. For this, the biosensor responses to the same substrate concentration were measured with an 12.5 min interval over one working day. The biosensor was characterized by relatively high signal reproducibility, standard deviation 2.93%.
Research on sensitivity of developed biosensors to heavy metal ions
Heavy metal ions can inhibit the biological activity of urease through the interaction with sulfhydryl groups of the enzyme active site:
Enz-Ser-OH + Me+ ^ Enz-Ser-O-Me + H+.
Fig. 1. Dependence of biosensor response from urea concentration in solution.
Measurement was performed in 5 mM phosphate buffer, pH 6.5
The enzyme inhibition by heavy metal ions causes a decrease in the amount of ions of ammonium and bicarbonate formed as a result of enzymatic reaction of the substrate conversion, and thus — a decrease of the biosensor response [19-21].
Using selected optimal concentration of urea, it was investigated the level of inhibition of urease biosensor by various heavy metals of the same concentration (Fig. 2). Initially, the solutions of nitrates of divalent metals zinc, copper, mercury and cadmium of the same concentration (100 pM) were taken, so, the concentration of all ions was also equal. The results showed that inhibition depends on the kind of ion and its ability to inhibit the enzyme described by the inhibition constant [22, 23].
Next, sensitivity of the developed biosensor to different concentrations of heavy metals ions was studied taking copper ions as an example. The calibration curve of the residual activity of urease-based bioselective element after inhibition (Fig. 3) showed its strong dependence on the concentration of copper ions in the measurment environment.
The stability of the biosensor inhibition after its incubation in the solution by heavy metal ions was also investigated (Fig. 4). For this, the responses of a number of biosensors to copper ions of the same concentration were measured after incubation under identical conditions. It was shown that the developed biosensors were characterized by relatively high signal reproducibility after inhibition with standard deviation of 2.6%.
Cd2‘ Zrf* Cu2* Hg2+ Ions
Fig. 2. Residual activity of urease biosensor after inhibition by different heavy metal ions.
Ions concentration — 100 pM.
Measurement was performed in 5 mM phosphate buffer, pH 6.5; * P <0.05 in comparison with initial activity of biosensor
18
Experimental articles
Fig. 3. Residual activity of urease biosensor after inhibition by Cu2+ ions with different concentrations. Inhibition time — 20 minutes.
Measurement was performed in 5 mM phosphate buffer, pH 6.5
Fig. 4. Reproducibility of urease biosensor after inhibitory analysis. Inhibitor concentration — 50 pM Cu2+. Inhibition time — 20 minutes.
Measurement was performed in 5 mM phosphate buffer, рН 6.5
Research and optimization of conditions of biosensor reactivation using EDTA
The urease inhibition by heavy metal ions is known to be irreversible, which is a cause of inefficiency of urease-based biosensors due to its one-time usage. Therefore, it was proposed to use a stage of biosensor reactivation for its repetitive employ, namely, to explore the possibility of reactivation of the developed biosensor after its inactivation by heavy metal ions. The initial urease activity after inhibition can be recovered by the reactivator EDTA, which displaces the heavy metal ion bound to serine residue in the enzyme. Thus the enzyme is reactivated, i.e. its ability to interact with the substrate is restored [14, 24].
Enz-Ser-OMe + EDTA ^ Enz-Ser-OH + EDT [Me]
The reactivation efficiency depends strongly on the time of reactivation, reactivator concentration, and other parameters. Therefore, the conditions of reactivation by EDTA should be first optimized. For this purpose, it was chosen an optimal time of biosensor incubation in the reactivator solution, which corresponds to the highest possible level of recovery of the enzyme membrane activity after inhibition. The optimal time of reactivation by 10 mM EDTA was 30 min.
It was also studied an efficiency of reactivation of urease biosensor after its inhibition during constant time (30 min) in the solution of copper ions of different
concentrations. Fig. 5 shows that reactivation under selected optimum conditions provided complete restoration of the enzymatic activity of membrane inhibited in the solution of copper nitrate of concentrations up to 400 pM.
Determination of number of cycles of reactivation of urease-based biosensor
At the next stage, the possibility of multiple reactivation of the developed biosensor was estimated with a purpose of its repetitive application for the inhibition analysis of heavy metal ions. In this case the biosensor can be used several times when working with irreversible inhibitors.
О
л 90 ->
Ъ • ■
s 604)
<= . ■
40> ■
о
л
§ 20£
GS o-|----,---1----,---1----p----,---p----1----p----1
0 250 500 750 1000 1250
. 2+
Concentration Си , цМ
Fig. 5. Dependence of biosensor reactivation level after inhibition by Cu2+ ions with different concentrations.
Measurement was performed in 5 mM phosphate buffer, рН 6.5
19
BIOTECHNOLOGIA ACTA, V. 8, No 6, 2015
The experimental results are shown in Fig. 6. First, the basic responses to urea were received (the first column). Next, the biosensors were incubated in a copper ions solution to obtain a certain residual activity of the enzyme (the second column). Further step was the biosensor reactivation in EDTA solution up to restoration of the urease activity to its original level (the third column). This procedure described was repeated until the bioselective membrane activity was completely restored.
The next cycles of inhibition and reactivation did not lead to complete restoration of the enzyme activity, but the sensitivity of biosensor was sufficient to reuse it for the analysis of heavy metal ions in the solution.
In the work, the urease-based biosensor is optimized for determination of heavy metal ions. The biosensor sensitivity was tested, the calibration curves were plotted for direct urea determination and for inhibition analysis of
heavy metal ions. The developed biosensor was characterized by high signal reproducibility at determination of both substrate and inhibitor. The possibility of biosensor reactivation after inhibition by heavy metal ions was proved. The optimum conditions of the process of reactivation using EDTA were selected.
It was established that the initial level of enzyme inhibition affects the capability of biosensor to be reactivated. It is shown that application of the reactivation stage allows repetitive usage of urease-based biosensors for inhibition analysis of heavy metal ions, which indicates the suitability of the method for multiple application.
The authors are grateful for financial support from the National Academy of Sciences of Ukraine in the frame of complex scientific and technical program “Sensor devices for medical-environmental and industrial-technological needs: metrological support and pilot exploitation”.
Fig. 6. Reproducibility application of urease based conductometric biosensor after inhibition by copper ions and reactivation by EDTA. Inhibition time in copper (II) nitrate solution (150 pM) — 20 min, reactivation time in EDTA solution (10 mM) — 30 min.
Measurement was performed in 5 mM phosphate buffer, рН 6.5; * P <0.05 in comparison
with initial activity of biosensor
20
Experimental articles
REFERENCES
1. Korobkin V. I., Peredelskiy L. V. Ecology. 12-th ed., add. and rec. Rostov n/D: Feniks. 2007, 602 p. (In Russian).
2. Evdokimova G.A. Microbiological activity of soils contaminated by heavy metals. Pochvovcdenie. 1982, V. 6, P. 125-132. (In Russian)
3. Alekseev Yu. V. Heavy metals in soils and plants. Leningrad: Agropromizdat. 1987, 142 p. (In Russian)
4. Vetoshkin A. G. Processes and mechanisms of hydrosphere protection. Penza: Izd-vo Penz. Gos. Un-ta. 2004, 188 p. (In Russian).
5. Kabata-Pendias A. Soil-plant transfer of trace elements — an environmental issue. Geoderma. 2004, 122, 143-149. doi:10.1016/j. geoderma.2004.01.004.
6. Song J., Zhao F. J., Luo Y. M., McGrath S. P., Zhang H. Copper uptake by Elsholtzia splendens and Silene vulgaris and assessment of copper phytoavailability in contaminated soils. Environmental Pollution. 2004, V. 128, P. 307-315. doi:10.1016/j.envpol.2003.09.019
7. Semenova A. D. Guidance for chemical analysis of surface waters. Leningrad: Gidrometeoizdat. 1977, 541 p. (In Russian)
8. Yatsenko V. Determining the characteristics of water pollutants by neural sensors and pattern recognition methods. Journal of Chromatography A. 1996, V. 722, P. 233-243. doi: 10.1016/0021-9673(95)00571-4.
9. Lee S-M., Lee W-Y. Determination of Heavy Metal Ions Using Conductometric Biosensor Based on Sol-Gel-Immobilized Urease. Bull. Korean Chem. Soc. 2002, 23(8), 1169-1171.
10. Anderson J. L., Bowden E. F., Pickup P. G. Dynamic electrochemistry: methodology and application. Anal. Chem. 1996, 68(12), 379-444. doi: 10.1021/a1960015y.
11. Tran-Minh C. Immobilized enzyme probes for determining inhibitors. Ion-Selective Electrode Rev. 1985, V. 7, P. 41-75.
12. Krajewska B. Urease Immobilized on Chitosan Membrane. Inactivation by Heavy Metal Ions. J. Chem. Tech. Biotechnol. 1991, 52(2), 157-162. doi: 10.1002/jctb.280520203.
13. Krajewska B., Zdarta J., Jesinowski T. Enzyme immobilization by adsorption: a review. J. Adsorption. 2014, 20(5), 801-821. doi: 10.1007/s10450-014-9623-y.
14. Zhylyak G. A., Dzyadevich S. V., Korpan Y. I., Soldatkin A. P., El’skaya A. V. Application of urease conductometric biosensor for heavy-metal ion determination. Sensors and Acuators B. 1995, V. 24-25, P. 145-148. doi:10.1016/0925-4005(95)85031-7.
15. Dziadevych S. V., Soldatkin O. P. Scientific and technologic foundations for miniature electrochemical biosensors development. Kyiv: Naukova dumka. 2006, 255 p. (In Ukrainian).
16. Schuvailo O. M., Danyleyko L. V., Arkhypo-va V. M., Dzyadevych S. V., El'skaya A. V., Cespuglio R., Soldatkin A. P. Development of microbiosensors based on carbon fibres for in vivo determination of glucose, acetylcholine and choline. Biopolym. Cell. 2002, 18(6), 489-495. (In Russian).
17. Soldatkin O. O., Kucherenko I. S., Pyeshko-va V. M., Kukla A. L., Jaffrezic-Renault N., El’skaya A. V., Dzyadevych S. V., Soldatkin A. P. Novel conductometric biosensor based on three-enzyme system for selective determination of heavy metal ions. Bioelectrochemistry. 2012, V. 83, P. 25-30. doi:10.1016/j.bioelechem.2011.08.001.
18. Soldatkin O. O., Kucherenko I. S., Marchenko S. V., Kasap B. O., Akata B., Soldatkin A. P., Dzyadevych S. V. Application of enzyme/ zeolite sensor for urea analysis in serum. Materials Science and Engineering: C, 2014, V. 42, P. 155-160. doi: 10.1016/j. msec.2014.05.028.
19. Kukla A. L., Soldatkin O. O., Pavluchen-ko A. S., Kucherenko I. S., Peshkova V. M., Arkhypova V. N., Dzyadevych S. V., Toxicity analysis of real water samples of different origin with ISFET multibiosensor. Biochemistry and Biophysics. 2014, 2(1), 7-12.
20. Gabrovska K., Godjevargova T. Optimum immobilization of urease on modified acrylonitrile copolymer membranes: Inactivation by heavy metal ions. Journal of Molecular Catalysis B: Enzymatic, 2009, 60(1-2), 69-75. doi:10.1016/j. molcatb.2009.03.018.
21. Zaborska W., Krajewska B., Olech Z. Heavy Metal Ions Inhibition of Jack Bean Urease: Potential for Rapid Contaminant Probing. Journal of Enzyme Inhibition and Medicinal Chemistry. 2004, 19 (1), 65-69. doi: 10.1080/1475636031000165023.
22. Zaborska W., Krajewska B., Leszko M., Olech Z. Inhibition of urease by Ni2+ ions: Analysis of reaction progress curves. Journal of Molecular Catalysis B: Enzymatic. 2001, 13(4-6), 103-108. doi: 10.1016/S1381-1177(00)00234-4.
23. Krajewska B., Zaborska W., Chudy M. Multi-step analysis of Hg2+ ion inhibition of jack bean urease. Journal of Inorganic Biochemistry. 2004, 98(6), 1160-1168. doi:10.1016/j.jinorgbio.2004.03.014.
24. Smolinska B., Cedzynska K. EDTA and
urease effects on Hg accumulation by Lepidium sativum. Chemosphere. 2007, 69(9), 1388-1395. doi: 10.1016/j.
chemosphere.2007.05.003.
21
BIOTECHNOLOGIA ACTA, V. 8, No 6, 2015
ДОСЛ1ДЖЕННЯ ТА ОПТИМ1ЗАЦ1Я РЕАКТИВАЦП УРЕАЗНОГО Б1ОСЕНСОРА ЗА 1НГ1Б1ТОРНОГО АНАЛ1ЗУ 1ОН1В ВАЖКИХ МЕТАЛ1В
Буднков С. Р.1, 2 Солдаткш О. О.1, 2
Кукла О. Л.3
Хоменко I. I.1, 4
Дзядевич С. В.1, 2 Солдаткш О. П.1, 2
Институт молекулярно! бмлоги i генетики НАН Укра!ни, Ки!в 2Ки!вський нацiональний унiверситет iMeHi Тараса Шевченка, Укра!на 31нститут фiзики напiвпровiдникiв iM. В. С. Лашкарьова НАН Укра!ни, Ки!в 4Центр штелектуально! власностi та передачi технологiй НАН Укра!ни, Ки!в
Метою роботи було визначення умов опти-мiзащl роботи уреазного б^сенсора пiд час аналiзу важких металiв та можливостi його реактивацп. Як кондуктометричний перетво-рювач використовували диференцшну пару золотих гребiнчастих електродiв, нанесених на керамiчну пiдкладку. Роль б^селективно! мембрани вiдiгравала уреаза, коiммобiлiзова-на з бичачим сироватковим альбумшом за до-помогою поперечного зшивання глутаровим альдепдом на поверхнi кондуктометрично-го перетворювача. Для iнгiбiторного аналiзу iонiв важких металiв пiдiбрано оптимальну концентращю субстрату — 1,0 мМ сечовини. Перевiрено чутливiсть бiосенсора до рiзних iонiв важких металiв та побудовано калiбру-вальнi кривi. Показано, що запропонований бiосенсор характеризувався високою вщтво-рюванiстю сигналiв до та тсля процесу шпбу-вання з похибкою вимiрювання менше 3%. Доведено можлив^ть реактивацп б^селективно! мембрани розчином етилендiамiнтетраоцтовоl кислоти пiсля незворотного шпбування уреази важкими металами. ПЩбрано оптимальнi умо-ви реактивацп бiосенсорiв, зокрема залежнiсть рiвня реактивацп ввд часу та концентрацп iонiв важких металiв у розчинi. Встановлено, що, використовуючи додатковий етап реактивацп бiосенсорiв пiсля iнгiбування, можна значно тдвищити селективнiсть процедури бмсенсор-ного визначення дешв важких металiв.
Ключовi слова: кондуктометричний перетво-рювач, бiосенсор, уреаза, важш метали, iнгiбiторний аналiз, реактиващя ензиму.
ИССЛЕДОВАНИЕ И ОПТИМИЗАЦИЯ РЕАКТИВАЦИИ УРЕАЗНОГО БИОСЕНСОРА ПРИ ИНГИБИТОРНОМ АНАЛИЗЕ ИОНОВ ТЯЖЕЛЫХ МЕТАЛЛОВ
Будников С. Р.1, 2
Солдаткин А. А.1, 2 Кукла А. Л3
Хоменко И. И.1, 4
Дзядевич С. В.1, 2
Солдаткин О. П.1, 2
1Институт молекулярной биологии и генетики НАН Украины, Киев 2Киевский национальный университет имени Тараса Шевченко, Украина 3Институт физики полупроводников им. В. Е. Лашкарева НАН Украины, Киев 4Центр интеллектуальной собственности и передачи технологий НАН Украины, Киев
Целью работы было определение условий оптимизации работы уреазного биосенсора при анализе тяжелых металлов и возможности его реактивации. Как кондуктометрический преобразователь использовалась дифференциальная пара золотых гребенчатых электродов, нанесенных на керамическую подложку. Роль биоселективной мембраны играла уреаза, ко-иммобилизованная с бычьим сывороточным альбумином с помощью поперечной сшивки гутаровым альдегидом на поверхности кондуктометрического преобразователя. Подобрана оптимальная концентрация субстрата для ингибиторного анализа ионов тяжелых металлов — 1,0 мМ мочевины. Проверена чувствительность биосенсора к разным ионам тяжелых металлов и построены калибровочные кривые. Показано, что предложенный биосенсор характеризовался высокой воспроизводимостью сигналов до и после ингибирования с погрешностью измерения менее 3%. Доказана возможность реактивации биоселективных мембран раствором этилендиаминтетрауксусной кислоты после необратимого ингибирования уреазы тяжелыми металлами. Подобраны оптимальные условия реактивации биосенсоров, в частности зависимость уровня реактивации от времени и концентрации ионов тяжелых металлов в растворе. Установлено, что, используя дополнительный этап реактивации биосенсоров после ингибирования, можно значительно повысить селективность процедуры биосенсорного определения ионов тяжелых металлов.
Ключевые слова: кондуктометрический преобразователь, биосенсор, уреаза, тяжелые металлы, ингибиторный анализ, реактивация энзима.
22